Description
The Biological Safety Cabinet YR05270-1 // YR05271-1 is a critical tool for laboratories dealing with pathogens. Built with advanced CLEANFLOW™ technology, it provides clean, stable airflow powered by a high-performance, energy-efficient German EBM fan. The arched air intake reduces turbulence, ensuring quiet operation while protecting both the operator and the work environment. Additionally, long-lasting HEPA and ULPA filters achieve over 99.995% efficiency, filtering particles as small as 0.12 microns.
Market Price
The price of the YR05270-1 // YR05271-1 ranges between $6700 and $7000, depending on the model and specifications. This reflects the high-quality materials, such as 304 stainless steel, and the advanced airflow control and filtration technology. Investing in this cabinet ensures compliance with strict biosafety standards, safeguarding both personnel and samples.
Frequently Asked Questions
- Is this cabinet difficult to operate?
No, the cabinet is easy to use, featuring an LCD LED control panel. The BIOCASION™ system monitors pressure, airflow, and filter life in real time, alerting you to any failures or abnormal conditions. - What protection does this cabinet offer?
The cabinet protects against cross-contamination through an air curtain isolation system and efficient HEPA filters. - What is the size of the work area?
The work area measures 670×965×560 mm for the YR05270-1 and 670×1260×560 mm for the YR05271-1, providing ample space for various lab applications.
Advantages and Disadvantages
Key advantages include robust stainless steel construction, an efficient HEPA filtration system, and an ergonomic design that reduces operator fatigue. A potential drawback could be its size and weight, which may require careful planning in smaller labs.
Field Use
This cabinet is ideal for biomedical, pharmaceutical, and research labs handling hazardous biological agents. It maintains a sterile environment to prevent cross-contamination and ensures sample integrity. Its sloped front design and optimized lighting improve operator comfort during long work shifts, making it indispensable in high-demand environments.
Recommendations
For optimal performance, regularly inspect and replace the HEPA filters, and calibrate the airflow system. It’s also crucial to check the alarm and UV protection systems to ensure safety and continuous operation.
Technical Parameters
Model | YR05270-1 | YR05271-1 |
Clean Rating | International standard ISO14644.1 CLASS 5(US209E,CLASS 100) | |
Inflow Velocity | 0.53m/s | |
Down Flow Velocity | 0.35m/s | |
Down Flow Velocity Precision | ±0.015m/s | |
Noise Level(dB) | ≤65 | |
Vibration | ≤3μm(X、Y、Z direction) | |
HEPA | Efficiency 99.995% | |
Light Intensity(LX) | ≥650 | |
Pressure Difference Range (Pa) | 0~500 | |
Operate Environment | environment temp.10-30℃,relative humidity under 70%, no obvious vibration and dust | |
External Dimensions (H*W*D) (mm) | 2120×1035×740 | 2120×1340×740 |
Dimension of Working Zone (H*W*D) (mm) | 670×965×560 | 670×1260×560 |
Power Supply | ||
Power(W) | 300 | 350 |
Net Weight (KG) | 175 | 210 |
Gross Weight (KG) | 195 | 235 |
Package Size (H*W*D) (mm) | 1710×1205×910 | 1710×1510×910 |