Description
The YR06407 Electrolyte Analyzer is an advanced tool that offers an interactive man-machine menu, real-time dynamic display of sample identifiers, and real-time system diagnostics. It also detects and filters small bubbles to prevent blockages and ensure accurate measurements. Its wave theory and direct pipe flushing methods prevent blockages and cross-contamination.
This device performs automatic calibration and two-point correction, offering quality control charts and statistical parameters printouts. Power failure protection prevents data loss, and it supports fuzzy queries, LIS software, and selectable data upload formats. Additionally, it manages the RTC clock and can be turned off at any time to reduce reagent consumption, making it ideal for various hospitals.
Market Price
The price range for an electrolyte analyzer varies depending on the specifications and supplier. Generally, these devices can be found between $2,000 and $10,000 USD. For an accurate and personalized quote, we recommend using the Kalstein Plus platform, where you can generate an automatic quote tailored to your specific needs.
Frequently Asked Questions
- What types of samples can this device analyze?
- The analyzer can work with serum, plasma, whole blood, and diluted urine.
- How long does it take to perform a measurement?
- The measurement speed is less than 30 seconds, allowing for efficient workflow in clinical settings.
- What is the storage capacity?
- It can store up to 10,000 test results, facilitating the management of large volumes of data.
Advantages and Disadvantages
Advantages:
- Intuitive interface and real-time diagnostics.
- Automatic calibration and correction.
- Prevention of blockages and cross-contamination.
- Power failure protection and efficient reagent management.
Disadvantages:
- Dependence on a constant power source.
- Possible need for regular maintenance to ensure accuracy.
Product Use in the Field
The YR06407 Electrolyte Analyzer is used in hospitals and laboratories to measure electrolyte concentrations in various biological samples. Its robust design and ability to prevent blockages ensure reliable performance even in high-volume environments. Additionally, its quick result acquisition makes it indispensable in medical emergency situations.
Recommendations
To maximize the use and maintenance of the Electrolyte Analyzer, it is recommended to perform regular calibrations and follow the manufacturer’s instructions for cleaning and storage. It is also beneficial to use specific cleaning solutions to keep the electrodes in optimal condition.
The YR06407 Electrolyte Analyzer is an essential investment for any institution seeking precision and efficiency in clinical analysis.
Request a quote today through our Kalstein Plus platform and discover how this device can transform your daily operations. Don’t wait any longer and ensure the best quality in your results!
Technical specifications
Model | YR06407 | YR06408 | YR06409 | YR06410 | YR06411 |
Sample | serum, plasma, whole blood and dilute urine | ||||
Storage | up to 10000 test results | ||||
Printer | Internal thermal printer | ||||
Interface | RS232 port | ||||
Analysis Method | ion selective electrode (ISE) | ||||
Measuring speed | <30s | ||||
Sample position | 39 positions (including 5 emergency and 4 functional positions) | ||||
Temperature | 5-40°C | ||||
Relative Humidity | <80% | ||||
Atmospheric pressure | 86~106 kPa | ||||
Power supply | AC220V±22V, 50Hz±1Hz | ||||
Power | <120W | ||||
Test Parameters | K Na Cl | K Na Cl TCO AG | K Na Cl iCa nCa TCa pH | K Na Cl iCa nCa TCa pH TCO AG | K Na Cl Li |
Model | YR06412 | YR06413 | YR06414 | YR06415 | YR06416 | YR06417 |
Sample | serum, plasma, whole blood and dilute urine | |||||
Storage | up to 10000 test results | |||||
Printer | Internal thermal printer | |||||
Interface | RS232 port | |||||
Analysis Method | ion selective electrode (ISE) | |||||
Measuring speed | <30s | |||||
Sample position | 39 positions (including 5 emergency and 4 functional positions) | |||||
Temperature | 5-40°C | |||||
Relative Humidity | <80% | |||||
Atmospheric pressure | 86~106 kPa | |||||
Power supply | AC220V±22V, 50Hz±1Hz | |||||
Power | <120W | |||||
Test Parameters | K Na Cl iCa nCa TCa pH Li | K Na Cl iCa nCa TCa pH Li TCO AG | K Na Cl iCa nCa TCa pH Mg | K Na Cl iCa nCa TCa pH Mg TCO AG | K Na Cl iCa nCa TCa pH Mg Li TCO AG |
K Na Cl iCa nCa TCa pH Mg Li
|
Accessories
Model | Description |
YRA264 | K Electrode |
YRA265 | K Electrode |
YRA266 | Cl Electrode |
YRA267 | Ca Electrode |
YRA268 | pH Electrode |
YRA269 | Li Electrode |
YRA270 | Mg Electrode |
YRA271 | Ref. Electrode |
YRA272 | Reagent Pack Calibration A 430ml; Calibration B 110ml; ( For Model H900 only) |
YRA273 | Reference Electrode Internal Solution – 10ml |
YRA274 | Electrode Internal
Solution – 3ml |
YRA275 | Activating Solution – 110ml |
YRA276 | QC Solution Level 2
(Water-based) – 110ml |
YRA277 | Cleaning Solution – 110ml |
YRA278 | Electrode Wash Solution (Protein Enzyme) -125mg (25mg/bottle x5) |
YRA278-1 | Electrode Wash Solution (Diluent) – 5ml ( 1 bottle) |